MiniSim Model 1000
Multi - Parameter Patient
SimulatorThe MiniSim 1000 is a
powerful, comprehensive patient simulator with a number of unique features. It provides a
full 12 lead ECG signal with more than 40 arrhythmias, dual blood pressure, respiration,
and temperature. It also delivers square, sine, and a triangle performance waveforms.
MiniSim 1000 arrhythmia selections
include pacer waveforms, ST segment changes, and both atrial and ventricular waveforms.
The dual blood pressure and respiration waveforms are synchronized to the ECG and track
with all arrhythmia selections. Multiple static and dynamic pressure readings are
selectable as per apnea events
Operation of the hand - held MiniSim 1000 is
simple via menu driven tactile keys utilizing a large LCD Display. The instrument may be
either AC or Battery operated.
Testing Features:
| 12 Lead ECG
|
| Over 40 Arrhythmias
|
| ECG Performance Waveforms
|
| Dual Blood Pressure
|
| Respiration
|
| Temperature
|
Operating Features:
| Easy to Use
|
| Small, Compact Size
|
| Menu Driven Operation
|
| Large LCD Display
|
| Tactile Keypad
|
| Battery or AC Operation
|
Specifications:
ECG
Normal Sinus:
Rate: |
30, 60, 70, 80, 90, 100,120, 150, 180, 210,
240, 270, 300, 350 BPM |
Amplitude: |
0.15 to 5mV |
High Level Output: |
0.5 Volt/mV of Low Level (Lead II) |
Performance Waveforms:
Sine, Triangle, Square
Rate: |
0.1 to 100 Hz |
|
0.1 to 1 Hz in 0.1 Hz increments |
|
1 to 10 Hz in 1 Hz increments |
|
10 to 100 Hz in 10 Hz increments |
Tall T Wave
Test Waveform@ 70 BPM
Atrial Pacer
Ventricular Pacer*
Paced Rhythms - Variable from 0.1 to 2mS
R Wave detection - Variable form 10 to 120mS.
Pulse:
2mS Pulse of 1mV amp @ 4 sec intervals
Pacers
ST Segment Analysis:
ST Elevation and Depression
ST Depression Positive and Negative Slope
Myocardial Infarction
Arrhythmias:
Atrial
Ventricular |
|
Blocks Respiration:
Normal
Physiological Simulation: |
Base Line Impedance: |
250, 500, 750, 100 Ohms |
Delta Impedance: |
0.1, 0.5, 1.0, 1.5 Ohms |
Rates: |
15, 30, 60, 120 BPM |
Apnea: |
Off, Continuous, 12 Seconds
and 32 Seconds Apnea-Bradycardia |
Accuracy: |
Delta Impedance 5% of
selection |
Rate: |
1% of selection |
Blood Pressure:
Impedance: |
350 Ohms |
Excitation: |
2 to 16 Volts |
Sensitivity: |
5uV/V/mmHg |
Static: |
0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, 300 mmHg. |
Dynamic: |
50/10, 60/20, 70/30, 80/40, 100/60, 120/80 mmHg. |
Accuracy: |
(±2% FS. ±1 mmHg |
Synchronized with all normal sinus rates |
Physiologically track all arrhythmia
selections. |
Temperature:
Compatibility: YSI 400/700
25/37 Degree Centigrade
Accuracy: ±1% of reading
General:
Display: |
16 Character 2 Line LCD |
Key Pad: |
8 tactile keys Ð F1, F2, F3, F4, >, <, ON/OFF, Menu |
Output Connectors ECG: |
10 Banana Jacks |
High Level: |
3.5mm Phono Jack |
Pressure: |
5 contact circular Din |
Temperature: |
3.5mm Stereo Jack |
Physical Size: |
5 1/2 x 31/2 x 1 1/2 inches |
Case: |
High Impact Plastic |
Weight: |
10 Oz. |
Power:
9 Volt Alkaline/AC adapter (Optional).
Battery Life 40 Hours Continuous.
Display Indicates Low Battery.
|
Model 1000 Software Waveform Selections
Atrial: |
Version |
1. Sinus Arrhythmia |
1.0 |
2. Missing beat @ 80 bpm |
1.0 |
3. Atrial Flutter |
1.0 |
4. Atrial Fibrillation |
1.0 |
5. Paroxysmal Atrial Tachycardia |
1.3 |
6. Junctional Premature Contraction |
1.3 |
Ventricular: |
|
1. One PVC Per Key Stroke |
1.0 |
2. Run of 3 PVCÕs |
1.0 |
3. Run of 6 PVCÕs |
1.0 |
4. 12 PVCÕs/min |
1.0 |
5. 24 PVCÕs/min |
1.0 |
6. Continuous PVCÕs |
1.0 |
7. Bigeminy |
1.0 |
8. Trigeminy |
1.0 |
9. Ventricular Flutter |
1.0 |
10. Ventricular Tachyardia |
1.0 |
11. Ventricular Fibrillation |
1.0 |
12. Right Focal PVC |
1.3 |
Tall
T Wave |
|
1. Tall T wave Rejection |
1.3 |
Automatic
Test Sequence |
|
1. R wave detection |
1.3 |
2. Tall T wave rejection |
1.3 |
3. Pacemaker Pulse Rejection |
1.3 |
4. Time to alarm for Tachycardia |
1.3 |
|
|
ST Segments: |
Version |
1. ST Elevation (ele) (ele: +slope, -slope, flat) |
1.2 |
2. ST Depression (dep) (dep: + slope, -slope, flat) |
1.2 |
3. Myocardial Infarction
| Ischemia (Isc)
|
| Injury (Inj)
|
| Infarction (Inf.)
|
| Inferior Infarction (Iinf)
|
|
1.2 |
Blocks:
|
|
1. First Degree AV Block |
1.3 |
2. Second degree AV Block: Mobitz I |
1.3 |
3. Second degree AV Block: Mobitz II |
1.3 |
4. Third degree AV Block |
1.3 |
5. Right Bundle Branch Block |
1.3 |
6. Left Bundle Branch Block |
1.3 |
7. Left Anterior Hemiblock |
1.3 |
Pacers: |
|
1. Runaway Pacer |
1.3 |
2. Atrial Pacer |
1.3 |
3. Ventricular Pacer |
1.3 |
4. Atrial Ventricular Pacer |
1.3 |
5. Asynchronous Pacer |
1.3 |
|
MiniSim Ordering Information |
Model 1000 |
Cat. No. |
Description |
300-(version) |
MiniSim 1000 with software indicated |
301 |
Soft Carrying Case |
302 |
AC Adapter-110 VAC |
302-220 |
AC Adapter-220 VAC |
Pressure Cables |
303 |
Pressure Cable Unterminated |
304 |
Temperature Cable |
305 |
Spacelabs Dual |
305-S |
Spacelabs Single |
306 |
Datascope Dual |
307 |
HP Dual |
308 |
HP Single |
309 |
Fukuda Dual |
310 |
Datascope Single |
311 |
Bard Medical Dual |
311-S |
Bard Medical Single |
312 |
Mennen Medical |
313-B |
Marquette (Rectangular) Eagle Series |
314 |
Ivy Biomedical |
316 |
Datex Single |
1000 |
ECG Snap to Banana Plug Adapter |
Additional Cables Available |
Model 330 |
330 |
MiniSim 330 EEG Simulator |
301 |
Soft Carrying Case |
|