DELTA 3000
Defibrillator /
Pacemaker Analyzer / ECG Arrhythmia Simulator |
|||
|
|||
Overview: The
Delta 3000 combines the complete testing functions of a defibrillator tester, a
pacemaker analyzer, and an ECG arrhythmia simulator into one compact, lightweight, and
easy to use instrument. No other instrument delivers the performance testing power of the Delta 3000. It will test all defibrillators, all external pacemakers including transcutaneous and transvenous and all ECG equipment. The built-in user selectable test loads eliminate the need to carry extra modules to test the transcutaneous pacers.
|
|||
Highlights: |
|
||
¨ Compact, Lightweight and Rugged. ¨ Simple, Menu Driven Operation. ¨ Large, Easy to Read Display. ¨ Simultaneous Display of All Test Measurements. |
¨ Dedicated Print key for Immediate Documentation. ¨ RS 232 Interface. ¨ Battery or AC Operation. ¨ Two-Year Warranty. |
||
Test Functions: |
|
||
|
|||
Defibrillator Tester: Tests All Defibrillators - Semi -Automatic, Automatic, and those with
Bi-Phasic Waveforms. ¨ Measures: Energy, Peak Voltage, Peak
Current, and Cardio Sync. Time. ¨ Storage and Playback of Output
Waveforms. |
|||
Pacemaker Analyzer: Tests All External Pacemakers Transthoracic and Transvenous Pacers.
¨ Measures: Amplitude, Rate, Pulse
Width, Pulse Energy, Sensed and Paced Refractory Periods.
¨ 50 and 60 Hz Immunity Tests.
¨ Transthoracic Test Load: Fixed 50 Ohm
or 12 User Selectable Loads to 1000-Ohms.
¨ A / V Pacer Test Load: 500 Ohms.
|
|||
ECG Simulator: Tests All ECG Equipment. ¨ 12 lead ECG. User selectable Rates and Amplitudes. ¨ Performance Waveforms: Sine, Square,
Triangle & Pulse ¨ 12 Arrthymia Waveforms |
|||
Specifications: |
|
|||||
|
||||||
Energy Measurement: Load: 50 Ohms +/- 1% Range (High): 900 Joules, Res. 1 J. Resolution: 1 Joules Range (Low): 0 100 Resolution: 0.1 Joules. Accuracy: +/- 2% of the Reading Plus 1J for Hi Range or Plus 0.5J for Low Range. Maximum voltage: 5200 Maximum Current: 100 Amp (H)
Stored: Discharge waveform expanded 200:1 and
is repeated 8 seconds. It can be viewed through the paddles and through the ECG High Level
Output. |
¨Transcutaneous: Test Load: 50
Ohms (Fixed) and Variable 100
to 1000 Ohms user selectable. Current: 1 to
200 mA Accuracy: +/- 5%
+. 5 mA Pulse Width: 0.5 to 80 ms. Accuracy: +/- 1% +.1 ms. Refractory: 20 500 ms Accuracy: +/- 2 % or +10ms
¨ Atrial and Ventricular:
Test Load: 500 Ohms +/-1%
Current: 1 to 25 mA
Accuracy: +/- 5% +. 5 mA.
Pulse Rate: 30 to 800 ppm.
Pulse Width: 0.5 to 80 ms.
Accuracy: +/- 1% +.1 ms.
Refractory: 20 500 ms
Accuracy: +/- 2 % or +10ms Immunity Test: 50/60 Hz |
ECG Arrhythmia Simulator: ECG: Normal Sinus RhythmRate: 30, 60, 90, 120, 240 BPM Performance: Sine, Square, Triangle and Pulse Rate: 0.5 1, 2, 10, 15, 20, 25, 40 Hz. Arrhythmias: VFBC, VFBF, VTAC, AFIB PVC, PVC1, PVC2, BGY, RBB AFIB, ATRFT, NSR.
Waveform
Output: 12 Lead ECG and Paddle and HI level
Output Jacks.
Power Requirements: Two 9-Volt Alkaline Battery or AC adapter. Dimensions: 10 x 11 x 2.25 inches (L x W x H) (25.4 x 27.9 x 5.7 cm) Weight: 4 lbs (1.9 Kg) |
||||
|
|
|
||||
Accessories: |
|
Standard:AC Adapter, Universal Paddle Adapter, User Manual, Shrouded Banana Plugs (set of 2), Certificate of Calibration. |
Optional: Carrying Case, Paddle Adapters for various manufacturers. Please call Customer Service for part numbers. |